Installation and Interfacing
Model 3585A
2-8
h. Adjust the front-panel INTENSITY control to obtain the desired intensity on the CRT
screen. Adjust the FOCUS and ASTIG controls as follows:
1 . Set the FOCUS control to midrange.
2. Adjust the ASTIG (Astigmatism) control for the sharpest trace possible.
3 . Adjust the FOCUS control for the sharpest and clearest trace possible.
4. Repeat Steps 2 and 3 until optimum adjustment is obtained. If, after several itera
tions a sharp, clear presentation cannot be obtained, internal adjustments are pro
bably required. These adjustments must be performed by a qualified service techni
cian.
i. The CRT display should now appear as shown in Figure 2-4.
Verify that the Zero Response is present and is aligned with the first vertical line on the left
hand side of the CRT graticule.
REF -25. 0 dBm
MARKER 20 000 000. 0 Hz
10 dB/DIV
RANGE -25 0 dB..
-97.
1 dB ..
- - - -
- - - -
- - - - -
- - - -
- - - - � - - - -
- - - - -
- - - -
- - - -
- - - -
CENTER 20 000 000. 0 Hz
SPAN 4 0 000 000. 0 Hz
RBW 30 KHz
VBW 30 KHz
ST
•
2 SEC
Figure 2·4. Turn·On Display
j . Press the front-panel
��t;�
key. (This will force an internal verification test and
Automatic Calibration. The "CALIBRATING" message will appear on the CRT screen.)
If the beeper sounds and/or a Calibration Error Code (e.g. , "CALIBRATION ERROR
03 ") appears on the CRT screen, the instrument is either defective or in need of adjustment.
Turn the instrument off and see the Fault Isolation information in Volume Two or the
Preliminary Troubleshooting procedures in Volume Three.
k . The 3585A's specifications are met after a 20-minute warmup at the ambient operating
temperature.
NOTE
When the internal Oven Reference is enabled (about ten minutes
after turn on), the beeper will sound and the "LOCAL OSC.
UNLOCKED " message will momentarily appear on the CR T
screen.
Summary of Contents for 3585A
Page 264: ......