Model 3585A
Performance Tests
o. Repeat Steps g thru i .
p . Disconnect the Audio Oscillator . This completes the Harmonic Distortion Test .
h p 3 585A
SPECTRUM ANAL Y Z E R
1iilC:J
j
jjjj
S Y N T HE S I Z ER
� �
0
� � �
� �
�
---1 --1 --1 --1 --1 --1
� ---1 --1 --1 --1 --1 --1 --1
�
�
---1 --1 --1 --1 --1 --1 --1 ���
� : :
� ---1 --1
--1 --1 --1 --1
rn
e o
o 0 0 0
J I
9MHz
J2
LOW PASS
F I LTER
( SEE FIGURE 4-21 )
Figure 4·4. Harmonic Distortion Test
4·26. Intermodulation Distortion Test
This test places two signals 1 00 Hz apart at the input of the 3585A. The second and third
order IM products are then checked against the specification.
Specification:
Intermodulation Distortion:
For two signals, each at least 6 dB below the
RANGE setting and separated in frequency by at
least 100 Hz, referred to the larger in frequency of
the two signals .
<
-80 dB; except 2nd order IM with one or both of
the input signals within the range of 10 MHz to 40
MHz,
<
-70 dB
Equipment Required:
Frequency Synthesizer . . . . . . . . . . . . . . . -hp- 3335A
Frequency Synthesizer . . . . . . . . . . . . . . . -hp- 3330B
1 0 dB/Step Attenuator . . . . . . . . . . . . . . . . -hp-355D
1 dB/Step Attenuator . . . . . . . . . . . . . . . . -hp- 355C
Frequency Summer . . . . . . . . . . . . (See Figure 4-20)
4-25