Model 3585A
Performance Tests
Table 4·6. Semi·Automatic Performance Test Equipment List
·hp- 0 3 5 8 5- 1 000 1 Semi-Automatic Performance Test Cartridge
* -hp- 9 8 2 5A Programmable Calculator
* -hp- 9 8 7 1 A Character I mpact Printer, Option 0 1
* -hp- 3 3 3 5A Frequency Synthesizer
* -hp- 3330B Frequency Synthesizer
* -hp- 3 4 5 5A Digital Voltmeter
* - hp- 9 8034A HP-IB I nterface
Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . . -hp- 5 3 2 8A
Function Generator . . . . . . . .
' . . . . . . . . . . . . . . . . . -hp- 3 3 1 1 A
Audio Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . -hp- 339A
500 Return Loss Bridge . . . . . . . . . . . . . . . . . . . . . . -hp- 8 7 2 1 A
7 50 Return Loss Bridge . . . . . . . . . . . . . . -hp- 8 7 2 1 A Option 008
500 Feed Thru Termination . . . . . . . . . . . . . . . . . . -hp- 1 1 048C
7 50 Feed Thru Termination . . . . . . . . . . . . . . . . . . . -hp- 1 1 094B
1 OdB/Step Attenuator . . . . . . . . . . . . . . . . . . . . . . . .
-hp- 3 5 5 0
1 dB/Step Attenuator . . . . . . . . . . . . . . . . . . . . . . . . . -hp- 3 5 5C
0 . 5V Thermal Voltage Converter . . . . . . . -hp- 1 1 0 5 1 A Option 0 1
Double Balanced M ixer . . . . . . . . . . . . . . . . . . . . . -hp- 1 0 5 34A
HP-IB Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -hp- 1 06 3 1
1 0: 1 Voltage Divider Terminated i n 500 . . . . . . . (See Figure 4- 1 1 )
Frequency Summer . . . . . . . . . . . . . . . . . . . . . (See Figure 4-20)
9 MHz Low Pass Filter . . . . . . . . . . . . . . . . . . . . (See Figure 4-2 1 )
DC Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . -hp- 6 2 1 3A
501 7 50 Minimum Loss Pad . . . . . . . . . . . . . . . . . . . -hp- 8 5 4 2 8 B
4·34. Test Equipment Substitutions
The included Semi-Automatic Performance Test software is designed to be used with the
calculator, printer and frequency synthesizers listed under Recommended Test Equipment,
Table 4-6. Other HP-IB compatible controllers and instruments may be used for the tests;
however, the user must write his own software to be compatible with his particular equip
ment. Substitute test equipment must meet the critical specifications listed in Table 4-1 .
NOTE
HP-IB is Hewlett-Packard's implementation of IEEE std
488-1975, "standard digital interface for programmable in
strumentation ".
4·3 5. Manual Tests
Before pro ceding with the calculator controlled portion of the semi-automatic performance
tests, two manual tests must be performed. The first of these tests is for frequency accuracy,
and the second test checks the 1 M ohm input impedance and capacitance.
4·36. Frequency Accuracy
This test verifies the frequency accuracy of the 3585A by using an external counter to check
the internal frequency reference. It is important that the frequency counter used to do this
test has a reference which is more accurate than that of the 3585A.
Specification: Counter Accuracy, ± O.3Hz ± lxlO-7/month