![HP 3585A Service Manual Download Page 78](http://html.mh-extra.com/html/hp/3585a/3585a_service-manual_161430078.webp)
Model 3585A
Performance Tests
Specification:
LO Feed Through
<
-15 dB below range
Procedure:
a. Disconnect all inputs to the 3585A.
b. Set the 3585A controls for:
INSTRUMENT PRESET
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 dBm
MANUAL SWEEP . . . . . . . . . . . . . . . . . . . . . . O Hz
c. Read the marker amplitude. The reading should be less than - 1 5 dB.
d. Record the marker amplitude reading under zero response on the Operational Verifica
tion Test Card.
4-22. Low Frequency Responses
Within the 3585A there are several frequencies which may be picked up by the sensitive
analog circuits. These frequencies include:
60 Hz· Power Line
5 kHz AID clock
20 kHz (approx) Power Supply Switching Oscillator
25 kHz (approx) CRT High Voltage Oscillator
100 kHz Fractional N Clock
1 MHz Fractional N Step Loop Clock
10 MHz Internal Reference
·or other power line frequency.
These frequencies and their harmonics will be used to verify that all Low Frequency
Responses are less than -120 dBm.
Specification:
Residual Responses
<
-120 dBm
Procedure:
a. Disconnect all inputs to the 3585A.
b. Set the 3585A controls for:
INSTRUMENT PRESET
REFERENCE LEVEL . . . . . . . . . . . . . . . . . -75 dBm
RES. BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Hz
VIDEO BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Hz
MANUAL SWEEP . . . . . . . . . . . . . . . . . . . . . . . . on
4-1 9