Model 3585A
Adjustments
1.
Adjust A 17R26 for a .oodB marker reading.
m. Enter RES BW Step
@]
on the 3585A keyboard.
n. Adjust A1 7R28 for a .oodB marker reading.
o. Enter RES BW Step
on the 3585A keyboard.
p. Adjust A 17R30 for a .000B marker reading.
q. Enter RES BW Step
on the 3585A keyboard.
r. Adjust A1 7R32 for a .oodB marker reading.
s. Enter RES BW Step
@]]
on the 3585A keyboard.
t. Adjust A17R34 for a .oodB marker reading.
5·32. 1 6dB Amplifier Adjustment
a. Disconnect the synthesizer from the A1711 connector.
b. Connect the Tracking Generator output to a lOdES/step attenuator. Connect the
lOdB/step attenuator to a 1 dB/step attenuator and place a SOD termination on the output of
the 1 dB/step attenuator. Connect the output of the termination to the A1711 connector.
c. Set the 3585A controls for:
RECALL 601
INSTRUMENT PRESET
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . 350KHz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . 100KHz
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10KHz
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2dB
MANUAL SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . " . . . . . -25dBm
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . -28dBm
d. Adjust the Tracking Generator amplitude for a marker amplitude reading of
-28.00dBm.
e. Set the 3585A controls for:
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
ENTER OFFSET
f. Set the external attenuators for 1 6dB of attenuation.
g. Set the 3585A REFERENCE LEVEL to -44dBm.
h. Adjust A1 8R77 for an offset marker amplitude reading of - 16.oodB.