Figure 5·43. 1 MO Input Capacitance Adju!itment Set·Up
c. Set the 3585A controls for:
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -25dBm
START FREQUENCy . . . . . . . . . . . . . . . . . . . . . . 100Hz
STOP FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . I MHz
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I dB
d. Using the Continuous entry Control, adjust the REF LVL so that the trace is centered
on the display.
e. Press the STORE A - B key on the 3585A.
f. Set the 3585A for a RANGE of -5dBm.
g. Adjust the attenuator for 20dB.
h. Adjust AIC18 so that the A trace overlaps the B trace as closely as possible (see Figure
5-44).
i. Remove all inputs to the 3585A.
5-67