Model 3585A
Adjustments
5-34.
INPUT SECTION
This section contains procedures to make the required adjustments on the Input board .
These adj ustments include Calibrator Symmetry, Input flatness , Autorange trip points,
Calibrator Output Level , I MO flatness , Amplitude and Input capacitance, LO Feedthrough
and Harmonic Distortion.
a. Being careful not to harm any of the cables connected to the Input/Conversion Sec
tion, turn the Input box on its side so that the bottom (Input Section , Al board) is accessable
(see Figure 5-36.).
Figure
5-36.
Input/Conversion Box Positioning For Adjustment
5-35.
Calibrator Symmetry Adjustment.
a. Set the 3585A controls for:
RECALL 605
INSTRUMENT PRESET
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -25dBm
AUTO RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . 20. I MHz
b . Remove the cable from A l l3 .
c. Slowly replace the cable until a display similar to figure 5-37 i s obtained . When this
display is obtained , do not push the cable in any further .
d . Adjust the CAL SYMMETRY control , R52, for the maximum possible marker
amplitude.