Adjustments
Model 3585A
5-10
f. Using a 10: 1 probe, connect the oscilloscope to the "OSC" test point (A67TP l ) .
WAR N I N G
The voltages involved in the following measurements may cause
serious injury or death. USE EXTREME CA UT/ON.
g. Turn the 3585A power on. Turn the front panel intensity control fully C. W . Verify
that A67TP l measures approximately 26Vp-p centered + 1 8V above ground potential.
Figure
5·5.
High Voltage Oscillator Output
h. Turn the front panel intensity control fully off and verify that the A67TP 1 output
is + 1 8V DC .
i. Disconnect the scope probe.
j . Connect a DVM (200V range) to the l 00V A67TP2. Adjust A67R38 for a reading of
l 00V
±
0.25V.
k. Turn the 3585A power off. Disconnect DVM.
WAR N I N G
The voltages present inside the high voltage power supply box can
cause serious injury or death.
Never
place an un insulated conduc
tive tool or object inside this box.
1.
Set the intensity control to the "9 o'clock" position .
m. Remove the aluminum cover from high voltage section (see Figure 5-6 for screw loca":'
tions) on the bottom side of the 3585A.