Probe Field
You use the Probe field to set the probe attenuation factor for the input
channel currently displayed in the Input field.
Probe Attenuation Factor
The probe attenuation factor can be set from 1:1 to 1000:1 in increments of
one. When you select the Probe field, the probe attenuation factor can be
changed by turning the knob or by entering a value from the pop-up keypad.
When you select a probe attenuation factor, the actual sensitivity at the input
does not change. The voltage values shown on the display (V/div, offset,
trigger level) are automatically adjusted to reflect the attenuation factor.
The marker and automatic measurement voltage values change when Run is
touched and the next acquisition is displayed
The default value for the Probe field is 10:1 for 10:1 divider probes.
Coupling Field
You use the Coupling field to set the input impedance for the channel
currently displayed in the Input field.
Coupling Field Selections
When you touch the Coupling field, a pop-up appears that shows the input
impedance values available for selection. The selectable values are as
follows:
•
1M
Ω
/ DC
•
1M
Ω
/ AC
•
50
Ω
/ DC.
The default value for the Coupling field is 1M
Ω
/ DC.
C A U T I O N
The maximum input voltage for the 50
Ω
/ DC Coupling field selection is 5
Vrms.
The Channel Menu
Probe Field
3-7
Summary of Contents for 16533A
Page 6: ...vi ...
Page 11: ...1 Run Stop Options ...
Page 15: ...2 Autoscale ...
Page 21: ...3 The Channel Menu ...
Page 23: ...Channel Field Menu Map The Channel Menu 3 3 ...
Page 29: ...4 Time Base Functions ...
Page 42: ...4 14 ...
Page 43: ...5 The Marker Menu ...
Page 45: ...Markers Field Menu Map The Marker Menu 5 3 ...
Page 46: ...Markers Field Menu Map Continued The Marker Menu 5 4 ...
Page 65: ...6 The Trigger Menu ...
Page 67: ...Trigger Field Menu Map The Trigger Menu 6 3 ...
Page 84: ...6 20 ...
Page 85: ...7 The Display Menu ...
Page 87: ...Display Field Menu Map The Display Menu 7 3 ...
Page 94: ...7 10 ...
Page 95: ...8 The Auto Measure Menu ...
Page 97: ...Auto Measure Menu Map The Auto Measure Menu 8 3 ...
Page 105: ...9 The Waveform Selection Menu ...
Page 107: ...Waveform Selection Menu Map The Waveform Selection Menu 9 3 ...
Page 116: ...9 12 ...
Page 117: ...10 Specifications and Characteristics ...
Page 125: ...11 Installation and Operator s Service ...
Page 140: ...11 16 ...
Page 148: ...Glossary 4 ...