2
Calibrate Vert Cal of the Self Cal.
a
Connect two BNC 50-
Ω
, 9-inch cables to the BNC tee adapter. Once
you touch Start, the instrument will prompt you to connect the cables
to the appropriate locations on the rear panel of the instrument.
b
Touch the Mode field, then touch Self Cal from the pop-up menu.
c
Touch the Procedure field, then touch Vert Cal from the pop-up menu.
d
Touch the Channel field, then touch All from the pop-up menu.
e
Touch the Start field and follow the instructions on the display.
f
After completion of Vertical Calibration, remove the cables from the
instrument.
3
Calibrate Delay of the Self Cal.
a
Obtain a BNC 50-
Ω
, 9-inch cable. Once you touch Start, the
instrument will prompt you to connect the cable to the appropriate
location on the rear panel of the instrument.
b
Touch the Procedure field, then touch Delay from the pop-up menu.
c
Touch the Channel field, then select channel 1.
d
Touch the Start field and follow the instructions on the display.
e
Repeat steps c and d for channel 2.
f
After completing all of the channel delay calibrations, remove the
cable from the oscilloscope.
4
If a multi-card HP 16533/34A module is being calibrated, repeat steps
1 through 3 for the next HP 16533/34A card until all cards have been
completed.
5
If a single-card HP 16533/34A module is being calibrated, proceed to
step 6. If a multi-card HP 16533/34A module is being calibrated,
perform the following steps before proceeding to step 6.
a
Remove power from the HP 16500 mainframe.
Installation and Operator’s Service
Self Cal menu calibrations
11–13
Summary of Contents for 16533A
Page 6: ...vi ...
Page 11: ...1 Run Stop Options ...
Page 15: ...2 Autoscale ...
Page 21: ...3 The Channel Menu ...
Page 23: ...Channel Field Menu Map The Channel Menu 3 3 ...
Page 29: ...4 Time Base Functions ...
Page 42: ...4 14 ...
Page 43: ...5 The Marker Menu ...
Page 45: ...Markers Field Menu Map The Marker Menu 5 3 ...
Page 46: ...Markers Field Menu Map Continued The Marker Menu 5 4 ...
Page 65: ...6 The Trigger Menu ...
Page 67: ...Trigger Field Menu Map The Trigger Menu 6 3 ...
Page 84: ...6 20 ...
Page 85: ...7 The Display Menu ...
Page 87: ...Display Field Menu Map The Display Menu 7 3 ...
Page 94: ...7 10 ...
Page 95: ...8 The Auto Measure Menu ...
Page 97: ...Auto Measure Menu Map The Auto Measure Menu 8 3 ...
Page 105: ...9 The Waveform Selection Menu ...
Page 107: ...Waveform Selection Menu Map The Waveform Selection Menu 9 3 ...
Page 116: ...9 12 ...
Page 117: ...10 Specifications and Characteristics ...
Page 125: ...11 Installation and Operator s Service ...
Page 140: ...11 16 ...
Page 148: ...Glossary 4 ...