Measurement Algorithms
Frequency (Freq)
The frequency of the first complete cycle displayed is measured using the
50% levels.
If the first edge on the display is rising, then
Freq =
1
t
rising
edge
2
−
t
rising
edge
1
.
If the first edge on the display is falling, then
Freq
=
1
t
falling
edge
2
−
t
falling
edge
1
.
Period
The period is measured at the 50% voltage level of the waveform.
If the first edge on the display is rising, then
Period
=
t
rising
edge
2
−
t
rising
edge
1
.
If the first edge on the display is falling, then
Period
=
t
falling
edge
2
−
t
falling
edge
1
.
Peak-to-Peak Voltage (Vp_p)
The maximum and minimum voltages for the selected source are measured:
Vp_p
=
V
maximum
−
V
minimum
,
where V
maximum
and V
minimum
are the maximum and minimum voltages
present on the selected source.
The Auto-Measure Menu
Measurement Algorithms
8-8
Summary of Contents for 16533A
Page 6: ...vi ...
Page 11: ...1 Run Stop Options ...
Page 15: ...2 Autoscale ...
Page 21: ...3 The Channel Menu ...
Page 23: ...Channel Field Menu Map The Channel Menu 3 3 ...
Page 29: ...4 Time Base Functions ...
Page 42: ...4 14 ...
Page 43: ...5 The Marker Menu ...
Page 45: ...Markers Field Menu Map The Marker Menu 5 3 ...
Page 46: ...Markers Field Menu Map Continued The Marker Menu 5 4 ...
Page 65: ...6 The Trigger Menu ...
Page 67: ...Trigger Field Menu Map The Trigger Menu 6 3 ...
Page 84: ...6 20 ...
Page 85: ...7 The Display Menu ...
Page 87: ...Display Field Menu Map The Display Menu 7 3 ...
Page 94: ...7 10 ...
Page 95: ...8 The Auto Measure Menu ...
Page 97: ...Auto Measure Menu Map The Auto Measure Menu 8 3 ...
Page 105: ...9 The Waveform Selection Menu ...
Page 107: ...Waveform Selection Menu Map The Waveform Selection Menu 9 3 ...
Page 116: ...9 12 ...
Page 117: ...10 Specifications and Characteristics ...
Page 125: ...11 Installation and Operator s Service ...
Page 140: ...11 16 ...
Page 148: ...Glossary 4 ...