Fall time
Fall time is measured between the 10% and 90% points of the falling edge.
To obtain the best possible measurement accuracy, set the sweep speed as
fast as possible while leaving the falling edge of the waveform on the display:
Fall time
=
t
10%
−
t
90%
.
Preshoot and Overshoot
Preshoot and Overshoot measure the perturbation on a waveform above or
below the top and base voltages (see the "Top and Base Voltages" section
earlier in this chapter). These measurements use all data displayed on the
screen; therefore, it is very important that only the data of interest be
displayed. If you want to measure preshoot and overshoot on one edge of a
waveform, then only display that edge. If you want to measure the maximum
preshoot and overshoot on a waveform, then display several cycles of the
waveform.
Preshoot is a perturbation before a rising or a falling edge and is measured
as a percentage of the top-base voltage.
Overshoot is a perturbation after a rising or a falling edge and it is measured
as a percentage of the top-base voltage.
If the measured edge is rising, then
Preshoot
= [
V
base
−
V
minimum
V
top
−
V
base
]
×
100
and,
Overshoot
= [
V
maximum
−
V
top
V
top
−
V
base
]
×
100.
If the measured edge is falling, then
Preshoot
= [
V
maximum
−
V
top
V
top
−
V
base
]
×
100
and,
Overshoot
= [
V
base
−
V
minimum
V
top
−
V
base
]
×
100.
The Auto-Measure Menu
Measurement Algorithms
8-10
Summary of Contents for 16533A
Page 6: ...vi ...
Page 11: ...1 Run Stop Options ...
Page 15: ...2 Autoscale ...
Page 21: ...3 The Channel Menu ...
Page 23: ...Channel Field Menu Map The Channel Menu 3 3 ...
Page 29: ...4 Time Base Functions ...
Page 42: ...4 14 ...
Page 43: ...5 The Marker Menu ...
Page 45: ...Markers Field Menu Map The Marker Menu 5 3 ...
Page 46: ...Markers Field Menu Map Continued The Marker Menu 5 4 ...
Page 65: ...6 The Trigger Menu ...
Page 67: ...Trigger Field Menu Map The Trigger Menu 6 3 ...
Page 84: ...6 20 ...
Page 85: ...7 The Display Menu ...
Page 87: ...Display Field Menu Map The Display Menu 7 3 ...
Page 94: ...7 10 ...
Page 95: ...8 The Auto Measure Menu ...
Page 97: ...Auto Measure Menu Map The Auto Measure Menu 8 3 ...
Page 105: ...9 The Waveform Selection Menu ...
Page 107: ...Waveform Selection Menu Map The Waveform Selection Menu 9 3 ...
Page 116: ...9 12 ...
Page 117: ...10 Specifications and Characteristics ...
Page 125: ...11 Installation and Operator s Service ...
Page 140: ...11 16 ...
Page 148: ...Glossary 4 ...