4
Seat the cards and tighten the thumbscrews. DO NOT use a Torx
screwdriver to tighten the thumbscrews.
W A R N I N G
For correct air circulation, filler panels must be installed in all unused card
slots. Correct air circulation keeps the instrument from overheating. Keep
any extra filler panels for future use.
5
To configure a multi-card module, use the master/expander trigger
cable included with the accessory kit of each HP 16533/34A card.
Starting with the top-most HP 16533/34A card, connect the ECL EXT
TRIG OUT to the ECL EXT TRIG IN of the card immediately below.
Repeat for all cards in the module.
Installation and Operator’s Service
To configure the HP 16533/34A module
11–7
Summary of Contents for 16533A
Page 6: ...vi ...
Page 11: ...1 Run Stop Options ...
Page 15: ...2 Autoscale ...
Page 21: ...3 The Channel Menu ...
Page 23: ...Channel Field Menu Map The Channel Menu 3 3 ...
Page 29: ...4 Time Base Functions ...
Page 42: ...4 14 ...
Page 43: ...5 The Marker Menu ...
Page 45: ...Markers Field Menu Map The Marker Menu 5 3 ...
Page 46: ...Markers Field Menu Map Continued The Marker Menu 5 4 ...
Page 65: ...6 The Trigger Menu ...
Page 67: ...Trigger Field Menu Map The Trigger Menu 6 3 ...
Page 84: ...6 20 ...
Page 85: ...7 The Display Menu ...
Page 87: ...Display Field Menu Map The Display Menu 7 3 ...
Page 94: ...7 10 ...
Page 95: ...8 The Auto Measure Menu ...
Page 97: ...Auto Measure Menu Map The Auto Measure Menu 8 3 ...
Page 105: ...9 The Waveform Selection Menu ...
Page 107: ...Waveform Selection Menu Map The Waveform Selection Menu 9 3 ...
Page 116: ...9 12 ...
Page 117: ...10 Specifications and Characteristics ...
Page 125: ...11 Installation and Operator s Service ...
Page 140: ...11 16 ...
Page 148: ...Glossary 4 ...