Data Management
62
Execution confirmation for [Save Measurement History]
This is a confirmation message for saving the measurement history to a USB memory stick.
Fig. 62 Execution confirmation for [Save Measurement History]
The buttons and button functions, which can be used with the execution confirmation for [Save
Measurement History] appearing, are described in the table below.
Table 13 Button functions with execution confirmation
for [Save Measurement History]
If the ENT button is pressed, the measurement history is saved to a USB memory stick as a
file in CSV format.
Each line describes one set of data. The values are separated by commas (",").
Saved items and formats are as shown in the table below.
Output example:
Date,Value,Unit,Value(Raw),Unit(Raw),Status,Memo
2001/01/01 12:34:56,123,mg/g,123,mg/L,0,sample01
2001/01/02 12:34:56,123,mg/g,123,mg/L,2,sample02
ESC button
Returns to the Data Top screen
page 57
Button
Function
Page
Button
Function
Page
ENT button
Saves the measurement history to a USB memory stick.
-
ESC button
Closes the message.
-
Item (1st line is the title)
Format
Remarks
Date
yyyy/mm/dd hh:MM:ss
Date and time of measurement (year, month, day, hour,
minute, second)
Value
ddddd
Concentration after display conversion
Unit
uuuuu
Concentration units after display conversion (mg/L, mg/kg,
mg/g, mg/PC)
Value (Raw)
ddddd
Raw concentration value
Units (Raw)
uuuuu
Units of raw concentration value (fixed at mg/L)
Status
sssss
Error flag
0: No error 2: Warming up
Memo
mmmmmmmmmmmmmmm
Entered memo