Maintenance
105
11.
Set the measurement mode to "MANUAL" on the Measurement Setting screen.
Reference
"Measurement condition settings" (page 23)
"Measurement Setting screen" (page 70)
12.
Press the MEAS button or up/down button until the process display on the screen
shows [Manual Measurement].
13.
Use a measuring cylinder or measuring syringe (for solvent) to dispense 20 mL to
30 mL of clean solvent through the sample inlet.
14.
Press the ENT button.
The instantaneous value measurement screen appears.
15.
Close the drainage valve and open the fill cell valve referring to "Screen operations
in the manual measurement/calibration mode" (page 14).
Cell filling starts.
16.
Make sure that there is no liquid leakage from the periphery of the filter block. If
liquid is leaking, remove the filter block, check the condition of the packing and
filter element and then reinstall.
17.
Open the fill cell valve and drainage valve referring to "Screen operations in the
manual measurement/calibration mode" (page 14).
Liquid drainage starts.
18.
After liquid drainage is finished, press the ESC button.
The start state of the manual measurement mode returns.
19.
Close the right cover.