Product Information
119
Product Information
Specifications
List of optional parts
Model
OCMA-500
Product name
Oil content analyzer
Measurement method
Solvent extraction - non-dispersive infrared absorption analysis method
Measured objects
Substances extracted from sample water into solvent and having infrared
absorption near a wavelength from 3.4
m to 3.5
m
Measurement range
0 mg/L to 200 mg/L
Resolution
For mg/L 0 to 99.9: 0.1, 100 to 200: 1
For mg/g, mg/kg, mg/PC 0 to 9.99: 0.1, 10.0 to 99.9: 0.1, 100 to 200: 1
Repeatability
0 mg/L to 9.9 mg/L: ±0.2 mg/L ±1 dig.
10.0 mg/L to 99.9 mg/L: ±2.0 mg/L ±1 dig.
100 mg/L to 200 mg/L: ±4 mg/L ±1 dig. * For standard liquids
Display method
3.5 inches, 320×240 dots
Backlight Color graphic LCD
Calibration method
Zero, span calibration
Amount of test sample required
Sample water : Solvent = 2:1
Extraction solvent
S-316
Amount of extraction solvent
required
8 mL
Extraction method
Built-in extractor
Ambient operating temperature
0°C to 40°C (no condensation)
Power supply
AC 100 V to 240 V ±10%, 50/60 Hz
Power consumption
AC 100 V: Approx. 60 VA, AC 240 V: Approx. 90 VA
External dimensions
342 (H) × 200 (W) × 313 (D) mm
Mass
Approx. 7 kg
External output
Output to a USB memory stick
Functions
300-item data memory
Self error determination
Stabilized measurement value display
Clock
Name
Part No.
Specifications
Dropper
3011025237
Polyethylene, 2.5 mL
Microsyringe
3200043748
25
L
Measuring syringe
3014054647
For solvent, 20 mL
Measuring syringe
3014054648
For sample, 20 mL
Absorbent sheet
3200549145
3 in package
Filter element
3200043516
For water filter, diameter 40 mm, 5 in package
Packing
3200044264
For water filter
Packing
3200564367
For extractor
B-heavy oil
3200043747
10 mL
Solvent S-316
3200044490
1.5 kg