Setting
71
The buttons and button functions, which can be used with the Measurement Setting screen
appearing, are described in the table below.
Table 24 Button functions with the Measurement Setting screen
Sample Vol.
Sets the sample volume used as a coefficient when the
measured value is converted to mg/kg, mg/g, or mg/PC.
1.0 to 10000.0
(The units depend on
the set [Measurement
Unit].)
page 76
Zero Shift Value
Sets the shift correction value for zero liquid.
-100.0 to 100.0 (mg/L)
page 77
Use Light
Sets whether the LED that illuminates the inside of the
extraction tank is ON or OFF.
OFF, ON
page 77
Confirm Save
Sets whether measurement data is automatically saved to
internal memory after a measured value is settled or a
message appears for execution of the data save.
AUTO, MANUAL
page 78
Save Memo
Sets whether a memo indicating the measurement conditions
or other information is saved together with the measured
values, if measurement data is saved.
OFF, ON
page 78
Display Negative
Sets whether a negative value is displayed or zero is
displayed, if the measured value is a negative value.
OFF, ON
page 79
Display Raw Data
Sets whether or not the concentration prior to conversion is
displayed together with the converted value, if converted
values are displayed.
OFF, ON
page 79
Item
Description
Setting range (unit)
or selections
Page
Button
Function
Page
ENT button
Shows the screen for setting the selected item.
-
Up button
Selects the next item up.
-
Down button
Selects the next item down.
-
ESC button
Returns to the Setting Top screen.
page 69