Measurement
42
Procedure
1.
Set the measurement mode to "SEMI-AUTO" on the Measurement Setting screen.
Change other settings as needed.
Reference
"Measurement condition settings" (page 23)
"Measurement Setting screen" (page 70)
2.
Press the MEAS button or up/down button until the process display on the screen
shows [Semi-Auto Measurement].
The semi-automatic measurement mode is entered.
Fig. 40 Start of the semi-automatic measurement mode
3.
Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 mL of
clean solvent through the sample inlet.
4.
Use the dropper to add one drop of hydrochloric acid through the sample inlet.
5.
Use the measuring cylinder or measuring syringe (for sample) to dispense 16 mL of
sample water through the sample inlet.
6.
Press the ENT button.
Extraction starts. The screen title changes to [Semi-Auto Measurement] and the process
display shows [Extraction].
The operation guide display shows [ENT: Next ESC: Previous].
The elapsed time appears (counting up).
Fig. 41 Semi-automatic measurement - Extraction