Measurement
40
Note
During layer separation, make sure that the solvent and sample water separate sufficiently. If the
solvent and sample water are insufficiently separated, press the ESC button to cancel
measurement, and lengthen [Separation Time] in the measurement settings. If the nature of the
sample is such that layer separation is difficult, it is recommended that you use semi-automatic
measurement or manual measurement in advance to determine the optimum [Separation Time].
If the elapsed time exceeds 60 seconds and the measured value does not appear, replace the
water filter (refer to "Replacing the water filter" (page 99)).
Tip
If [Confirm Save] is set to "MANUAL", the operation guide display will show [ENT: Save ESC:
Skip]. To save the measured value, press the ENT button. To skip, press the ESC button. For
details on the [Confirm Save] setting, refer to "Confirm Save" (page 78).
If [Save Memo] is set to "ON" in the measurement settings, a pop-up screen to input the data
name will appear immediately before the measured value is saved. To save the measurement
conditions or other memo with the measured value, enter the data name and press the ENT
button. To save only the measured value and not an additional memo, press the ESC button. For
details on the [Save Memo] setting, refer to "Save Memo" (page 78). For details on the pop-up
screen to input the data name, refer to "Character keys" (page 18).
7.
Repeat steps 3. to 6. until the set [Number of Purge] is completed and measurement
starts.
Fig. 38 Auto Measurement - Measurement
After the result appears, the measurement liquid is drained. During drainage, the
process display on the screen shows [Drainage]. When drainage ends, the start
state of the automatic measurement mode returns.