Setting
80
Calibration Setting screen
If [Calibration Setting] is selected on the Setting Top screen, the Calibration Setting screen
opens.
The Calibration Setting screen shows the current calibration settings.
If an item is selected with the up/down button and the ENT button is pressed, a pop-up screen
appears to let you change the setting of the selected item.
Fig. 88 Calibration Setting screen
The setting items on the Calibration Setting screen are as shown in the table below.
Table 25 Items on the Calibration Setting screen
The buttons and button functions, which can be used with the Calibration Setting screen
appearing, are described in the table below.
Table 26 Button functions with the Calibration Setting screen
Item
Description
Setting range (units)
Page
Span Point
Sets the concentration of the span liquid for calibration.
1.0 to 200.0 (mg/L)
page 81
Extraction Time
Sets the extraction time in the automatic calibration mode.
0 or 10 to 600 (sec)
page 81
Separation Time
Sets the layer separation time in the automatic calibration
mode.
0 or 10 to 600 (sec)
page 82
Number of Purge
Sets the numbers of purges in the automatic calibration
mode.
0 to 9 (time)
page 82
Calib. Mode
Sets the calibration mode.
AUTO, MANUAL
page 83
Button
Function
Page
ENT button
Shows the screen for setting the selected item.
-
Up button
Selects the next item up.
-
Down button
Selects the next item down.
-
ESC button
Returns to the Setting Top screen.
page 69