Calibration
29
6.
Press the ENT button.
7.
Repeat steps 3. to 6. until the set [Number of Purge] is completed and measurement
starts.
If the set [Number of Purge] has not been completed:
The purge sequence starts. The screen title changes to [Auto Span Calibration
(Purge: 1)], and the process display changes as follows as the sequence
progresses:
[Extraction]
→
[Layer Separation]
→
[Fill Cell]
→
[Drainage]
The remaining time of the current process is shown (counting down).
When the purge sequence ends, the start state of the automatic span calibration
mode returns.
When the remaining [Number of Purge] is 0:
The measurement sequence starts. The screen title changes to [Auto Span
Calibration: Measurement] and the process display changes as follows as the
sequence progresses:
[Extraction]
→
[Layer Separation]
→
[Fill Cell]
→
[Calibration]
The remaining time of the current process is shown (counting down) during
extraction, layer separation, and cell filling. The elapsed time is shown (counting up)
during calibration. When calibration ends, the calibration result appears.
After the result appears, the measurement liquid is drained. During drainage, the
process display on the screen shows [Drainage].
Fig. 26 Display of automatic span calibration result
When drainage is completed, the start state of the automatic span calibration mode
returns.