Measurement
24
Meas mode
Moving average setup screen
*Setup menu
If "non" is selected from the "diSP" setting
items, nothing is displayed in the sub-display
part after returning to the Meas mode.
Controlled value setup menu
Cleaning check screen
Calibration data check screen
Setting
Change
Establishes
the value
Cancels the
value
Alarm settings can be changed during
measurement.
The upper or lower limits of motions cannot
be switched.
The delay time of an alarm contact point can
be changed.
When a numerical value is adjusted with the
▲
/
▼
key and the ENT key is pressed, the
changes will be confirmed.
If the use of cleaning contacts is enabled, the
countdown of the cleaning cycle timer can be
checked.
Cleaning contacts are output while the ENT
key is pressed in the test mode.
The calibration history can be checked.
The value of an amount of transmitted light obtained during the
latest zero calibration is compared with the factory-configured
value, and is expressed in %.
This screen shows the number of days elapsed since the last
zero calibration.
The moving average count can be changed during
measurement.
When a numerical value is adjusted with the
▲
/
▼
key and
the ENT key is pressed, the changes will be confirmed.
See page 26.
See page 26.
See page 26.