Installation
18
Detector installation method
Two types of holders for installing the detector have been standardized.
Immersion-type holder
The detector is maintained at the tip of a holder, and the holder is immersed in a sample.
Drop-type holder
The detector is passed through the inside of a holder fixed in place, and only the detector is
raised and lowered.
For detailed handling of each holder type, refer to the instruction manual that comes with the
product.
J
For an immersion type holder
Even if there are variations in the water surface levels of measured liquid, be sure to set up an
immersion type holder while the tip of a detector is always soaked in the measured liquid.
If an immersion type holder is set up at a location where measured liquid flows in only one
direction, set up and orient a detector in a downstream direction.
When an immersion type holder is used, remove the adaptor that is attached to the detector.
Active voltage contact is output from a converter to drive a solenoid valve for cleaning.
Installation example for
immersion-type holder
Cleaning water
20
0 mo
re
S
amp
le
liqu
id
leve
l