Measurement
31
Setting to enable
or disable the
use of cleaning
contact.
"no"
"yES"
"no"
Specify whether to output active voltage
contact for cleaning.
Type of cleaning
output
(when the use of
cleaning contact
is set to "yES")
"nor"
"or"
"And"
"trg"
or
Select a combinational operation
between the internal timer and the
external directive for cleaning.
"nor": This is the normal setting. An
external directive is not accepted and
cleaning contact is output in response
to only the cleaning cycle defined by
the internal timer. However, a
transmission signal is held by an
external directive.
"or": The cleaner operates in response
to the internal timer, and cleaning
contact is output while an internal
directive is input.
"And": Cleaning contact is output in
response to the internal timer only while
an external directive is input.
"trg": This disables the capabilities of
the internal timer. Cleaning contact is
output in response to only an external
directive (trigger for 2 s or longer)
during specified cleaning time.
After the end of cleaning, a Hold
contact is output according to the
settings for the extended time of
transmission output hold.
(
See “
Setting for
internal timer
cleaning cycle
(when a cleaning
output type is set
to "nor", "or", or
"And")
0.1 to
168.0 h
2
Define a cleaning cycle in units of
hours.
Setting for
internal timer
cleaning time
(when the use of
cleaning contact
is set to "yES")
2 to 600 s
10
Define the cleaning time in units of
seconds.
Setting for the
extended time of
transmission
signal hold after
the end of
cleaning
2 to 600 s
120
Define the time to be held after the end
of cleaning.
Note: If cleaning output is set to "no",
the output after the release of test
cleaning is held.
RS-485 address
setting
0 to 99
0
Specify an address for RS-485
communication.
Screen 1
Screen 2
Description
Setting
range
Initial
value
Remarks