Measurement
30
Setting for a
range of
transmission
output
(current range)
"2000"
"5000"
"10000"
"20000"
"oPt"
0 to
10000
Full-scale transmission output can be
selected. In the options, zero and a
point in the full scale can be arbitrarily
defined.
Setting a zero
point of
transmission
output (when the
current range is
set to "oPt")
0 to
22000
0
The zero point of transmission output
can be freely selected within the
Range.
Setting a span
point of
transmission
output (when the
current range is
set to "oPt")
0 to
22000
10000
A span point of transmission output can
be freely selected in the range.
Setting the
transmission
hold common to
transmission
output (current
hold).
"HoLd"
"PrES"
"HoLd"
Define transmission output during the
HOLD mode.
z
"HoLd": The last value is output
while being held.
z
"PrES": The preset value is output.
Setting hold
value (when the
transmission
hold is set to
"PrES")
0 to
22000
10000
The HOLD state will be maintained at
the pre-set value.A preset value can be
defined outside the transmission range.
Setting for Hold
or Burnout
common to
transmission
output
"nor"
"out.4"
"out.20"
"nor"
Specify the burnout of transmission
output for the occurrence of an
abnormality during measurement.
z
out.4: Outputs 3.8 mA at the time of
burnout.
z
out.20: Outputs 21 mA at the time of
burnout.
z
non: There is no burnout setting.
Screen 1
Screen 2
Description
Setting
range
Initial
value
Remarks