Operation
21
Q
Initial setup
The converter is shipped with initial settings from the factory.
If there is a need to change any initial settings, follow the instructions given in “ Entering the
Setup menu ” (page 27).
For the initial settings, see the table below.
To return to the Meas mode, hold down the ESC key until the HOLD indicator turns OFF.
Setting
menu items
Screen
display
Description
Initial
value
Reference
page
Parameter
Setting
"PArA"
"dF"
Set the moving average count (damping factor).
20
"A.rEt"
Set auto return.
"yES"
"rEt.t"
Set auto return time.
2
Sensor settings
"SEnSor”
"tyPE"
Selection of a working curve type (Type).
"PtnA"
Display
settings
"diSP"
"S.diSP" Set the auxiliary display part (sub-display part).
"non"
"rng.C"
Set out of measurement range hidden (range cut).
"yES"
Relay output
settings
Setting
"r1" "r2"
"1.tyPE" Set contact output types ("r1" and "r2").
"non"
"1.HorL" Set a control method (when the contact output type is "AL").
"Hi"
"1.SEt"
Set an alarm triggering value (when a contact output type is set to
"AL").
4000
"1.dt"
Set contact output delay time (when a contact output type is set to
"AL" or "HoLd").
0
"1.CLnt"
Set a cleaning output type (when a contact output type is set to
"CLn").
"CLn"
FAIL output
setting
"FAiL"
"F.SC"
Set the FAiL output for abnormality detection by self-check.
"yES"
"F.oF"
Set the FAIL output for the case where a measured value exceeds
the full-scale value.
"no"
"F.dt"
Set FAIL output delay time (in seconds).
0
Transmission
output settings
"Curr"
"C.rng"
Set a range of transmission output (Current Range).
10000
"rng.0"
Set a zero point of transmission output (when Current Range is set
to "oPt").
0
"rng.5"
Set a span point of transmission output (when the current range is
set to "oPt").
10000
"C.HoLd"
Set the transmission hold common to transmission output (current
hold).
"HoLd"
"PrES
Set a Hold value (when transmission hold is set to "PrES").
10000
"b.out"
Set “Hold” or “Burnout” common to transmission output.
"non"
Cleaning
output
"CLn"
"CLn"
Output for a cleaner (Cleaning)
"no"
"tyPE"
Timing of cleaning start (Type)
"or"
"CyCLE" Cleaning cycle (cycle time)
2.0
"CLn.t"
Cleaning time (cleaning time in seconds)
10
"Hold.t"
Transmission output hold time after the end of cleaning (Hold time
in hours)
120
Setting for
RS485
"rS"
"Addr"
Set RS-485 address.
0