Measurement
26
」
」
Cleaning
"non"
Setting
This screen shows
that cleaning output
is not used.
"- - - - " indicates
that a relay is not
used.
Setting
"yES"
Cleaning output
wait time screen
The time up to next
cleaning start is
shown in hours,
minutes, and
seconds.
10 h or longer are
shown in hours and
minutes.
Cleaning output
test operation
screen
Cleaning contacts
are output while the
ENT key is
pressed.
For test operation.
Transmission
output is in the hold
state at the hold
time during
cleaning operation
and after cleaning.
」
Screen for check of
an mount of light for
zero calibration
This screen
displays a ratio of
the factory-
configured amount
of transmitted light
for zero calibration
to the amount of
transmitted light for
the latest zero
calibration.
You can be made
aware of any
deterioration in the
LED light source.
Screen displaying
the number of days
elapsed since the
last zero calibration
Indicates the
number of elapsed
days after the last
calibration.
Time is counted up
every 2 h during
energizing. Time is
not measured
during power-off.
Moving average
count check screen
(dumping factor)
The moving
average count can
be checked and
changed.
Increase dF to
reduce the
variations in
measured values.
Decrease dF to
speed up the
response.
Screen 1
Screen 2
Condition
Description
Action
Remarks