Measurement
32
Cal menu
In the Cal menu, zero calibration and span calibration of the MLSS detector is preformed.
Since the transmission method is adopted, the zero point might change due to dirt, but span
sensitivity does not change.
It is recommend to check the zero point at regular intervals. For span calibration, obtain the
correlation between a measured value and an analyzed value in a certain period during which
concentration fluctuates, and enter a coefficient. For measurement of activated sludge, select
a working curve of MLSS. For any samples other than activated sludge, good correlation can
be obtained up to high concentration by selecting a working curve suitable for the properties of
a sample.
An operational check board is not a benchmark. Span calibration cannot be performed by
attaching an operational check board.
Entering the calibration menu
Calibration can be performed by one of two methods: hold down the CAL key during
measurement, or select the Calibration menu in the Hold mode.
This section describes the procedure for entering Cal mode from the Hold mode.
1.
In the Meas mode, hold down the HOLD key until the HOLD indicator lights up.
2.
Press the
▲
/
▼
key to display "CAL" on the display part for the measured value.
3.
Press the ENT key.
4.
Press the
▲
/
▼
key to display a desired item on the auxiliary display.
5.
Press the ENT key.
Coefficient
adjustment
Cal menu
Display of
various
calibration
Establishes
the value
Cancels
the value
Hold down the ESC
key to Meas mode.
or
Returns to hold down
the HOLD key.
Hold down the HOLD
key to go to the Hold
mode.
Meas mode
Hold mode
or