Measurement
40
Cleaning output
Cleaning output (CLN output) is output in accordance with the following conditions.
Cleaning
output
Yes/No
"CLn"
Cleaning
output
type
Internal
cleaning cycle
Cleaning:
○
Standby: -
External
contact input
ON:
○
OFF: -
Cleaning
output *1
ON:
◎
OFF: -
"CLn"
Contact
output *2
ON:
●
OFF: -
"HoLd"
Contact
output *3
ON:
●○
OFF: -
Transmission
output
"yES"
"nor"
○
○
◎
●
●
*4
○
-
◎
●
●
-
○
-
-
○
-
-
-
-
-
Measured
value
"or"
○
○
◎
●
●
*4
○
-
◎
●
●
-
○
◎
●
●
-
-
-
-
-
Measured
value
"And"
○
○
◎
●
●
*4
○
-
-
-
-
Measured
value
-
○
-
-
-
-
-
-
-
-
"trg"
No setting
ON for 2 s
◎
●
●
*4
-
-
-
-
Measured
value
"no"
No setting
No setting
○
-
-
○
*4
-
-
-
-
Measured
value
*1:
Cleaning output is output in accordance with a specified method when the settings for cleaning time of cleaning
outputî of the Setup menu is set to "yES".
*2:
The "CLn" contact output is enabled in the HOLD state during cleaning and after cleaning.
*3:
●
indicates a case where the "Hold" contact output is enabled in the HOLD state during cleaning and after
cleaning.
○
indicates a case where the "Hold" contact output is enabled when the external contact input is ON.
*4:
The transmission output is enabled in accordance with the transmission output hold settings in the Setup menu.