Troubleshooting
48
Measurements exceeding the measurable range
If an SS measured value is outside the measurement range (below
−
200 mg/L or above
20000 mg/L [above 22000 mg/L in the case of range cut]), the measured value is displayed as
blinking.
Take remedial action as instructed below.
Symptom
Possible cause
Remedial action
Blinking display
(out of the
measurable
range)
A concentration of the
sample is high.
If the measured value of zero water is not abnormal, this trouble is out of
measurement range. No corrective action is possible.
The concentration of the
sample is low.
z
For clean water measurement, there is a possibility that the value is
negative. Perform zero calibration again.
z
Immerse it in samples of high concentration. If indications are
negative, contact us.