FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Schedule of
Maintenance
8
8-3
The following definitions apply for the terms listed below when referenced in this Schedule of
Maintenance.
ACR Phantom-
American College of Radiology Accreditation Phantom.
Contrast:
Percent of difference between object intensity and background intensity.
The system must be maintained twice per annum
. More regular intervals for radiation
surveys and image tests can be conducted by the user in accordance with the instructions in
this manual. Refer to local government regulations and facility procedures for further
guidance and compliance.
A radiation survey must be conducted every time the system is relocated or damaged.
Cybersecurity
Any software or firmware updates that have been released since the last maintenance that
resolve a safety or security issue can be provided, as necessary.
The firmware is hard coded into the control board so if a firmware update is required, replace
the control panel PCB.
8.5.0
Definitions
8.6.0
Periodic
Maintenance
Summary of Contents for Faxitron CT
Page 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Page 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Page 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Page 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Page 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Page 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Page 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Page 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...