FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Schedule of
Maintenance
8
8-13
Section 8.7.0 Failure Criteria:
The Safety Interlocks and Indicators test fails if any of the following conditions are observed:
•
Green Power LED does not illuminate when system is powered on,
•
Red X-Ray On LED does not illuminate during X-ray exposures,
•
Audible Intermittent Beep does not sound during X-ray exposures,
•
Touch Screen Display does not
show yellow and black “X
-
Ray On”
message. Note: Top left corner of the
screen will also display “X
-
Ray On” in blue text
,
•
X-ray exposure
does not stop within 2 seconds after “Stop X
-
ray” button on the display monitor, the
”Stop” button on the Touch Screen, or the physical “Stop” button on
the control panel is pressed,
•
X-ray exposure does not stop immediately when the door is open, or
•
The Touch Screen does not display “Door Open” any time the chamber door is open
.
Summary of Contents for Faxitron CT
Page 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Page 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Page 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Page 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Page 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Page 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Page 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Page 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...