FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Camera
Calibration
3
3-5
Uniformity Calibration Sequence
The uniformity calibration is to ensure that density measurement of the system (Hounsfield
units) is calibrated correctly.
As per the previous calibration, go to the calibration page and click on Uniformity Calibration.
This time insert the clear Uniformity Calibration phantom by placing it on the CT pedestal.
The phantom has a location stud that will center the phantom on the pedestal. Once positioned
click Okay.
Check the box to confirm the Phantom is placed and proceed by clicking Start.
As before the calibration will begin taking a full 360-degree image of the phantom which takes
up to four minutes.
3.1.3
Uniformity
Calibration