FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Compliance
and Safety
6
6-2
Requirements | Faxitron CT Compliance
Emission Limits
Radiation emitted from the Faxitron CT unit does not exceed an exposure of 0.1 milli-
Roentgen (mR) per hour at any point 5 centimetres (cm) outside the external surface.
For countries requiring emission values measured at 10 cm from the external surfaces, the
radiation does not exceed 1.0
u
S/hr per hour at 10 cm.
Safety Interlocks
Faxitron CT system has two independent safety interlocks on the specimen door.
Start Button
Following interruption of X-Ray generation by the safety interlocks, the Start button on the
front panel will resume X-Ray generation when ready.
Password Control
The control system is protected with a Password that must be entered to generate X-Rays.
Stop Button
Along with the software and touch panel Stop button there is a physical Stop button on the
Faxitron CT front panel to terminate the generation of X-Rays. This is in addition to the safety
interlocks on the system doors.
X-Ray Generation Indicators
The Faxitron CT has an X-Ray On indicator light on the front panel, on the touch panel and
graphical user interface, plus it produces an audible tone when generating X-Rays.
Power Ratings
120VAC 60Hz or 230VAC 50Hz 700W Max.
Tested Standards
Testing carried out by Intertek, a US Nationally Recognized Testing Laboratory and
European Notified Body
UL 61010-1
3rd Edition, 2012-05
Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory
Use.
IEC 61326-1: 2012
Electrical equipment for measurement, control and laboratory use - EMC requirements.
IEC 61326-2-6: 2013
Electrical equipment for measurement, control and laboratory use - EMC requirements. Part
2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment
Complies with DHHS radiation performance standards, 21 CFR, Subchapter J
DICOM Version
Faxitron CT uses the latest DICOM standard:
ACR-NEMA Digital Imaging and Communications in Medicine (DICOM) v3.0, Final Draft,
Aug. 1993.
And complies with:
NEMA PS 3.1 - 3.20 (2016), Digital Imaging and Communications In Medicine Set
See DICOM Conformance Statement for more Information
6.1.0
Compliance
Requirements and
Safety Measure
6.2.0
DICOM Standard
Summary of Contents for Faxitron CT
Page 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Page 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Page 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Page 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Page 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Page 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Page 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Page 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...