FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Camera
Calibration
3
3-3
For the daily image calibration ensure the chamber is completely empty, including the foam
tray. Check the box to confirm the chamber is empty and proceed by clicking Start.
During warmup and calibration, X-Rays may be generated. During X-Ray generation, a
:Cancel”
button will be made visible to allow a user the option to stop X-Rays through software or touch
control panel.
Geometric Calibration Sequence
Geometric Calibration ensures the system is set up correctly to perform the highly precise
reconstruction of the CT images (3D). This can be performed at any time to ensure image
quality but must be conducted at least every 3 months.
If the 3D image quality degrades at any time, visible rings in the image for example, conduct a
geometric calibration.
Go to the main screen and click on the Calibrate button, as before, and then select Geometric
Calibration.
3.1.1
Image Calibration
Cont.
3.1.2
Geometric
Calibration
Summary of Contents for Faxitron CT
Page 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Page 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Page 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Page 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Page 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Page 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Page 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Page 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...