FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Faxitron CT
Quick Start
4
4-3
Basic Imaging Procedures
Warning:
Use approved containers or gloves in order to avoid contact with any potentially
hazardous samples.
Caution
– Use of controls or adjustments or performance of procedures other than those
specified herein may result in hazardous radiation exposure.
•
Place sample in a sample container provided. The specimen must fit completely within the
inner ring shown below.
•
To ensure the DICOM data is correct, place the Posterior face down on the
foam with the Inferior (Caudal) side facing the maker shown below with the
arrow.
•
Place the tray on the pedestal with the marker facing out, as shown below. You may need to
rotate the tray side to side slightly to get it to sit down securely. Check that it is not free to
rotate and is stable.
•
Close the chamber door.
•
Select the “Start Procedure” from
the main screen
Figure 4.1.2
Faxitron CT Consumable Tray
Figure 4.1.3
Select Procedure Screen
Inner Ring
Summary of Contents for Faxitron CT
Page 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Page 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Page 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Page 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Page 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Page 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Page 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Page 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...