3 2303-20 TC MODULE
53
1
2
3
4
5
6
7
8
9
APPX
IDX
3.5.2 Function Specifications
Execute the functions from the PC application via communications.
Monitoring function
Measured value recording function
Alarm judgment and recording function
3.5.3 General Specifications
This function outputs the current measured values (instantaneous values).
Measurements are recorded at a set recording interval.
Real-time manage-
ment
This is automatically set from the PC application at the start of recording.
Recording start
Immediate start/Reserved-time start
Recording end
Manual end/Reserved-time end
Operation when
memory is full
Memory full stop /Endless
Set the condition before the start of recording.
Recording interval
1/2/5/10/15/20/30 sec., 1/2/5/10/15/20/30/60 min.
Recording mode
• Instantaneous value
• MAX/MIN/AVE
• Instantaneous value + MAX/MIN/AVE
Total 3 modes
Set the mode before the start of recording.
Recorded data
One data set contains time and temperature/humidity information (2 channel each).
Data is scaled if scaling is ON.
Recording capacity
512 k bytes Flash memory
Quantity of recorded
data
• Instantaneous value recording mode: 30,000 data × 2 CH
• MAX/MIN/AVE recording mode: 15,000 data × 2 CH
• Instantaneous value + MAX/MIN/AVE recording mode: 12,000 data × 2 CH
Power outage
protection
After recovering from a power outage, the instrument automatically returns to the state
held before the outage.
Alarm judgment is made at every sampling, and the history will be recorded in flash memory if the measured
value recording function remains active.
Judgment method
Criterion threshold can be set to either Hi or Lo.
The instantaneous value at every sampling is judged (effective in any measurement mode).
Recorded data
One data set contains time, generation/reversion, CH and judgment threshold infor-
mation.
Clock accuracy
±100 ppm (Reference value at temperature from 0 to 50°C (32 to 122°F) without the
communications module)
Backup
Recorded data (saved in flash memory)
Data loss for up to 2 minutes before and after a power outage may occur.
Communication
interface
CAN bus
Maximum rated voltage
to earth
33 Vrms, 70 VDC (Total of input voltage)
Alarm output
Open collector: 30 VDC, 20 mA max.
Rated supply voltage 5 V±0.3 VDC
Maximum rated power 1.4 W
Dielectric strength
1.4 kVAC Between input and alarm output, Input/Output and CAN bus (50/60 Hz,
Response current 5 mA, one minutes).
Summary of Contents for 2300 Smart Site
Page 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Page 13: ...Contents XII...
Page 21: ...8 Service...
Page 87: ...74 4 2304 20 PULSE MODULE...
Page 105: ...92 5 2304 21 PULSE MODULE...
Page 253: ...240 12 2342 20 OUTPUT MODULE...
Page 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Page 285: ...272 15 2353 20 LAN MODULE...
Page 317: ...304 18 2362 20 DC POWER MODULE...
Page 345: ...332 20 2392 01 02 MODULE BASE...
Page 347: ...334 21 Maintenance and Service...
Page 348: ......
Page 349: ......
Page 350: ......
Page 351: ......
Page 352: ......
Page 353: ......
Page 354: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 358: ......
Page 359: ......
Page 360: ......
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......
Page 365: ......
Page 366: ......