104
6 2305-20 INSTRUMENTATION MODULE
6.4.2
Insulation of Internal Circuit
In the 2305-20, the input circuit and alarm output are insulated
from the CAN bus as shown in the block diagram below. (With-
stand voltage: 1.4 kVAC, 50/60 Hz, Response current: 5 mA, 1
minute)
The CH1 and CH2 terminals are not insulated from each other.
When connecting signals different in potential to these terminals, use
an additional measurement module or insulate the signals externally
before connection to the terminals. This will prevent module errors and
malfunction.
Alnalog
circuit
CH1
/ CH2
Me
asu
rem
en
t
In
put
Digital
circuit
Insu
latin
g
cir
cuit
Module-to-
module com-
munications
circuit
CA
N bu
s
Co
nne
cto
r
Mo
dul
e ba
se
Po
w
e
r
Insulation
withstand voltage
of 1.4 kVAC
Insulation
withstand voltage
of 1.4 kVAC
AL
A
R
M
A
larm
ou
tpu
t
Ground
The COM terminal of the alarm output terminal is used for both
CH1 and CH2.
Summary of Contents for 2300 Smart Site
Page 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Page 13: ...Contents XII...
Page 21: ...8 Service...
Page 87: ...74 4 2304 20 PULSE MODULE...
Page 105: ...92 5 2304 21 PULSE MODULE...
Page 253: ...240 12 2342 20 OUTPUT MODULE...
Page 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Page 285: ...272 15 2353 20 LAN MODULE...
Page 317: ...304 18 2362 20 DC POWER MODULE...
Page 345: ...332 20 2392 01 02 MODULE BASE...
Page 347: ...334 21 Maintenance and Service...
Page 348: ......
Page 349: ......
Page 350: ......
Page 351: ......
Page 352: ......
Page 353: ......
Page 354: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 358: ......
Page 359: ......
Page 360: ......
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......
Page 365: ......
Page 366: ......