9 2331-20 POWER METER MODULE
173
1
2
3
4
5
6
7
8
9
APPX
IDX
9.3.4 Connecting Alarm Output
The terminal blocks use M3.0 screws (at a tightening torque of 0.5 N•m).
We recommend that you use a round crimp connector (RAV 1.25-
3) for connection.
Open collector output (Rating: 30 VDC, 20 mA max.)
Design the wiring to prevent short-circuiting of the ALARM
and CLAMP SENSOR INPUT terminals, thus avoiding electric
shock and accidents due to short-circuiting.
• The ALARM terminals are electrically insulated from the U
INPUT terminals and CLAMP SENSOR INPUT terminals.
• To avoid the effects of external noise, design the wiring so that
cables connected to the ALARM terminals are separated from
those for measurement, as well as cables connected to the
CLAMP SENSOR INPUT terminals and voltage cables.
Summary of Contents for 2300 Smart Site
Page 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Page 13: ...Contents XII...
Page 21: ...8 Service...
Page 87: ...74 4 2304 20 PULSE MODULE...
Page 105: ...92 5 2304 21 PULSE MODULE...
Page 253: ...240 12 2342 20 OUTPUT MODULE...
Page 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Page 285: ...272 15 2353 20 LAN MODULE...
Page 317: ...304 18 2362 20 DC POWER MODULE...
Page 345: ...332 20 2392 01 02 MODULE BASE...
Page 347: ...334 21 Maintenance and Service...
Page 348: ......
Page 349: ......
Page 350: ......
Page 351: ......
Page 352: ......
Page 353: ......
Page 354: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 358: ......
Page 359: ......
Page 360: ......
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......
Page 365: ......
Page 366: ......