For the power supply output cable, use a cable with a conductor cross
section of 1.0 mm
2
or more and length of up to 1 meter (3.28 feet).
Example:
UL1007 AWG18/AWG16 (equivalent to 0.75 mm
2
/1.25 mm
2
) or
equivalent
300 V vinyl-cabtire cable 2-core, 1.0 mm
2
or more
The cable length shall not exceed 1 meter.
We recommend using a round solderless terminal for the connection.
Example:
RAV1.25-4, BT1.25-F4
20.2.3 Connecting the CAN Cable
Extending the internal bus by connecting CAN cables to the CAN
terminals allows up to 63 measurement modules to be connected
to a single communications module.
Use a CAN cable that complies with ISO 11898.
(M4.0)
at tightening torque
1.2 N•m
(M4.0)
Tightening torque:
1.2 N•m
Summary of Contents for 2300 Smart Site
Page 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Page 13: ...Contents XII...
Page 21: ...8 Service...
Page 87: ...74 4 2304 20 PULSE MODULE...
Page 105: ...92 5 2304 21 PULSE MODULE...
Page 253: ...240 12 2342 20 OUTPUT MODULE...
Page 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Page 285: ...272 15 2353 20 LAN MODULE...
Page 317: ...304 18 2362 20 DC POWER MODULE...
Page 345: ...332 20 2392 01 02 MODULE BASE...
Page 347: ...334 21 Maintenance and Service...
Page 348: ......
Page 349: ......
Page 350: ......
Page 351: ......
Page 352: ......
Page 353: ......
Page 354: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 358: ......
Page 359: ......
Page 360: ......
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......
Page 365: ......
Page 366: ......