8 2321-20 WAVEFORM MODULE
149
1
2
3
4
5
6
7
8
9
APPX
IDX
[Example 5]
Monitoring the power of a voltage dip
The following are the procedures to monitor the power of a voltage dip.
Procedures
1.
As the input voltage region of the 2321-20 is ±50 V, connect
the tip of the input terminal to a step-down transformer to
reduce the voltage to ±50 V.
2.
Set scaling.
3.
Set the window in of the analog, trigger filter.
Set the voltage level to a level that the trigger cannot fall
below, and the power frequency of the trigger filter to above
half-wave (above 10 ms for 50 Hz and above 8.4 ms for 60
Hz). The trigger will be activated when the voltage is above
half-wave and below the pre-set level.
Summary of Contents for 2300 Smart Site
Page 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Page 13: ...Contents XII...
Page 21: ...8 Service...
Page 87: ...74 4 2304 20 PULSE MODULE...
Page 105: ...92 5 2304 21 PULSE MODULE...
Page 253: ...240 12 2342 20 OUTPUT MODULE...
Page 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Page 285: ...272 15 2353 20 LAN MODULE...
Page 317: ...304 18 2362 20 DC POWER MODULE...
Page 345: ...332 20 2392 01 02 MODULE BASE...
Page 347: ...334 21 Maintenance and Service...
Page 348: ......
Page 349: ......
Page 350: ......
Page 351: ......
Page 352: ......
Page 353: ......
Page 354: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 358: ......
Page 359: ......
Page 360: ......
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......
Page 365: ......
Page 366: ......