4.2 AC Power Cord
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 8
CareAssist™ Bed Service Manual (man330)
SHOCK HAZARD:
An abnormally high leakage current is symptomatic of a degradation
within the AC power cable and the power supply. Personal injury could
result if the value of the leakage current is above 500 microamperes.
5. Check that the leakage current of the bed is compliant (see “Leakage
current” on page 6-5 in the Preventive Maintenance Schedule).
6. To make sure the CareAssist™ Bed operates correctly, do the “Function
Summary of Contents for CareAssist P1170
Page 2: ......
Page 6: ...Revisions Page iv CareAssist Bed Service Manual man330 NOTES ...
Page 14: ...Page xii CareAssist Bed Service Manual man330 NOTES ...
Page 141: ...Warranty Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 5 5 NOTES ...
Page 147: ...Base Frame Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 11 5 NOTES ...
Page 157: ...Head Siderail Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 21 5 NOTES ...
Page 165: ...CPR Cables Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 29 5 NOTES ...
Page 187: ...SideCom Communication System Chapter 3 ...
Page 188: ...Siderail Interface P C Board 72207 and 71514 Chapter 3 ...
Page 189: ...Caregiver Control Chapter 3 ...
Page 190: ...Left Caregiver Signal Conditioning Chapter 3 ...
Page 191: ...Bed Control Board Power and Phase Control Chapter 3 ...
Page 192: ...Bed Control Board Power Management Chapter 3 ...
Page 193: ...Bed Control Board Battery Management Chapter 3 ...
Page 194: ...Bed Control Board Clock and Divider Chapter 3 ...
Page 195: ...Bed Control Board Relay FET and Drivers Chapter 3 ...
Page 196: ...Bed Control Board PLD and Connector Chapter 3 ...
Page 197: ...Bed Control Board Motor Current Limiter Chapter 3 ...