![Hill-Rom CareAssist P1170 Service Manual Download Page 60](http://html1.mh-extra.com/html/hill-rom/careassist-p1170/careassist-p1170_service-manual_2134163060.webp)
2.7 Foot Section Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 18
CareAssist™ Bed Service Manual (man330)
2.7
Foot Section Malfunction
1. At least one of the other functions is working.
Yes
No
↓
→
2. Do the following to make sure the controls operate correctly.
a. Disconnect all the controls.
b. Connect connect a known good control.
3. The controls operate correctly.
Yes
No
↓
→
4. Connect one control at a time until the bad control is located. Replace the
affected control.
5. Test the function of the foot section motor intermediate switch.
The problem only occurs with the Dining Chair™ Position and knee
section functions.
Yes
No
↓
→
7. Do the following:
a. Remove the power supply cover.
b. Swap the connectors of the knee section and foot section motors (#5
brown, #2 green) on the power board.
c. Activate the CPR function.
The foot section function operates correctly, the knee section motor rod
extends to mid-travel.
Yes
No
↓
→
Replace the foot motor (refer to procedure 4.8 on page 4-24).
8. Test the function of the thigh section motor intermediate switch:
a. Connect the foot section motor cable to its socket (#2, green).
b. Either connect a new thigh section motor in extended position or using
an ohmmeter, check the state of the intermediate switch of the thigh
section motor. The motor is extended, the contacts 3 and 4 of the 4-pin
Summary of Contents for CareAssist P1170
Page 2: ......
Page 6: ...Revisions Page iv CareAssist Bed Service Manual man330 NOTES ...
Page 14: ...Page xii CareAssist Bed Service Manual man330 NOTES ...
Page 141: ...Warranty Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 5 5 NOTES ...
Page 147: ...Base Frame Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 11 5 NOTES ...
Page 157: ...Head Siderail Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 21 5 NOTES ...
Page 165: ...CPR Cables Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 29 5 NOTES ...
Page 187: ...SideCom Communication System Chapter 3 ...
Page 188: ...Siderail Interface P C Board 72207 and 71514 Chapter 3 ...
Page 189: ...Caregiver Control Chapter 3 ...
Page 190: ...Left Caregiver Signal Conditioning Chapter 3 ...
Page 191: ...Bed Control Board Power and Phase Control Chapter 3 ...
Page 192: ...Bed Control Board Power Management Chapter 3 ...
Page 193: ...Bed Control Board Battery Management Chapter 3 ...
Page 194: ...Bed Control Board Clock and Divider Chapter 3 ...
Page 195: ...Bed Control Board Relay FET and Drivers Chapter 3 ...
Page 196: ...Bed Control Board PLD and Connector Chapter 3 ...
Page 197: ...Bed Control Board Motor Current Limiter Chapter 3 ...