Table 1-8. Electromagnetic Emissions Guidance
Table 1-9. Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The CareAssist™ Bed model P1170 is intended for use in the electromagnetic environment specified below. The customer
or the user of the model P1170 should make sure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
RF emissions
CISPR 11
Group 1
The model P1170 uses RF energy only for its internal
functions. Therefore, its RF emissions are low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class A
The model P1170 is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not Applicable
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Not Applicable
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The CareAssist™ Bed model P1170 is intended for use in the electromagnetic environment specified below. The customer
or the user of the model P1170 should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Radiated RF
IEC 61000-4-3
3 Vrms
80 MHz to 2.5 GHz
3 Vrms
80 MHz to 2.5 GHz
Portable and mobile RF communica-
tions equipment should not be used
at close distances to the P1170 bed.
(See Note 2)
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV on Power Supply
Lines
± 1 kV on Input/ Output
Lines
± 2 kV on Power Supply
Lines
± 1 kV on Input/ Output
Lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV Differential Mode
(Line-Line)
± 2 kV Common Mode
(Line-Ground)
± 1 kV Differential Mode
(Line-Line)
± 2 kV Common Mode
(Line-Ground)
Mains power quality should be that
of a typical commercial or hospital
environment.
Summary of Contents for CareAssist P1170
Page 2: ......
Page 6: ...Revisions Page iv CareAssist Bed Service Manual man330 NOTES ...
Page 14: ...Page xii CareAssist Bed Service Manual man330 NOTES ...
Page 141: ...Warranty Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 5 5 NOTES ...
Page 147: ...Base Frame Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 11 5 NOTES ...
Page 157: ...Head Siderail Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 21 5 NOTES ...
Page 165: ...CPR Cables Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 29 5 NOTES ...
Page 187: ...SideCom Communication System Chapter 3 ...
Page 188: ...Siderail Interface P C Board 72207 and 71514 Chapter 3 ...
Page 189: ...Caregiver Control Chapter 3 ...
Page 190: ...Left Caregiver Signal Conditioning Chapter 3 ...
Page 191: ...Bed Control Board Power and Phase Control Chapter 3 ...
Page 192: ...Bed Control Board Power Management Chapter 3 ...
Page 193: ...Bed Control Board Battery Management Chapter 3 ...
Page 194: ...Bed Control Board Clock and Divider Chapter 3 ...
Page 195: ...Bed Control Board Relay FET and Drivers Chapter 3 ...
Page 196: ...Bed Control Board PLD and Connector Chapter 3 ...
Page 197: ...Bed Control Board Motor Current Limiter Chapter 3 ...