Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 6
CareAssist™ Bed Service Manual (man330)
15. Unplug the power from the power source, and test the functions checked in
The bed functions operate properly.
Yes
No
↓
→
16. Press the Knee Up control until it reaches the high position and then press
the Foot Up control.
The foot section rises to the high position without stopping or the audible
alarm sounding.
Yes
No
↓
→
17. Press the Foot Down control.
The foot section drops to the low position without stopping or the audible
alarm sounding.
Yes
No
↓
→
18. Press the Knee Down control until knee section is in the horizontal
position. Press the Dining Chair™ Position control.
The head section and knee section rise until the required position is
reached.
Yes
No
↓
→
19. Press the Bed Flat control.
The head section, knee section and foot section return to the horizontal
position.
Yes
No
↓
→
20. Press the Bed Up control until the intermediate position is reached, and
then operate the Trendelenburg control.
The sleep surface gradually tilts to maximum Trendelenburg without any
problem or abnormal noise.
Yes
No
↓
→
21. Press the Reverse Trendelenburg control.
Summary of Contents for CareAssist P1170
Page 2: ......
Page 6: ...Revisions Page iv CareAssist Bed Service Manual man330 NOTES ...
Page 14: ...Page xii CareAssist Bed Service Manual man330 NOTES ...
Page 141: ...Warranty Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 5 5 NOTES ...
Page 147: ...Base Frame Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 11 5 NOTES ...
Page 157: ...Head Siderail Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 21 5 NOTES ...
Page 165: ...CPR Cables Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 29 5 NOTES ...
Page 187: ...SideCom Communication System Chapter 3 ...
Page 188: ...Siderail Interface P C Board 72207 and 71514 Chapter 3 ...
Page 189: ...Caregiver Control Chapter 3 ...
Page 190: ...Left Caregiver Signal Conditioning Chapter 3 ...
Page 191: ...Bed Control Board Power and Phase Control Chapter 3 ...
Page 192: ...Bed Control Board Power Management Chapter 3 ...
Page 193: ...Bed Control Board Battery Management Chapter 3 ...
Page 194: ...Bed Control Board Clock and Divider Chapter 3 ...
Page 195: ...Bed Control Board Relay FET and Drivers Chapter 3 ...
Page 196: ...Bed Control Board PLD and Connector Chapter 3 ...
Page 197: ...Bed Control Board Motor Current Limiter Chapter 3 ...