Electrical System
Chapter 3: Theory of Operation
Page 3 - 8
CareAssist™ Bed Service Manual (man330)
Patient Pendant
The patient pendant controls operate at 3.3 V DC supplied by the main bed
control and integrate three main principles:
• Filtering and amplification of inputs for electrostatic protection and signal
formatting
• Multiplexing for the management of the controls
• Demultiplexing for the control of the indicators
Data from the controls and indicators are transmitted in series (link type - SPI)
so as to reduce the number of wires. The clock signal that starts these
exchanges is controlled by the main bed control P.C board. All the data is sent
in tens of milliseconds. When the system is in stand-by mode, the power is cut
off and the clock signal is no longer controlled. The battery control is therefore
directly wired so as to have continuous operation.
Motor Characteristics
For motor characteristics, see table 3-4 on page 3-8.
Table 3-4. Motor Characteristics and Dimensions
Management of the Motors
The motors control signals are supplied by the main bed control P.C. board.
Their extreme positions are detected in series by upper and lower internal EOT
devices. These devices directly cut off the power supply for the head section,
knee section and foot section actuators. In addition, the head section and foot
section motors have intermediate internal microswitches to control the
automatic contour and Dining Chair™ position.
Description
Hilow column
motor
Head motor
Knee motor
Foot motor
Power
96 W (x2)
120 W
84 W
84 W
Voltage
18 to 40 V DC
18 to 40 V DC
18 to 40 V DC
18 to 40 V DC
Intermittent
service
10%
3 min/30 min
10%
2 min/18 min
10%
6 min/60 min
10%
6 min/60 min
Actuator/
column travel
400 mm
± 5 mm (15.75")
239 mm
± 2 mm (9.4")
60 mm
± 2 mm (2.36")
45 mm
± 2 mm (1.77")
Retracted rod/
center distance
445 mm
± 2mm (17.5")
310 mm
± 2mm (12.2")
325 mm
± 2mm (12.8")
Summary of Contents for CareAssist P1170
Page 2: ......
Page 6: ...Revisions Page iv CareAssist Bed Service Manual man330 NOTES ...
Page 14: ...Page xii CareAssist Bed Service Manual man330 NOTES ...
Page 141: ...Warranty Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 5 5 NOTES ...
Page 147: ...Base Frame Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 11 5 NOTES ...
Page 157: ...Head Siderail Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 21 5 NOTES ...
Page 165: ...CPR Cables Chapter 5 Parts List CareAssist Bed Service Manual man330 Page 5 29 5 NOTES ...
Page 187: ...SideCom Communication System Chapter 3 ...
Page 188: ...Siderail Interface P C Board 72207 and 71514 Chapter 3 ...
Page 189: ...Caregiver Control Chapter 3 ...
Page 190: ...Left Caregiver Signal Conditioning Chapter 3 ...
Page 191: ...Bed Control Board Power and Phase Control Chapter 3 ...
Page 192: ...Bed Control Board Power Management Chapter 3 ...
Page 193: ...Bed Control Board Battery Management Chapter 3 ...
Page 194: ...Bed Control Board Clock and Divider Chapter 3 ...
Page 195: ...Bed Control Board Relay FET and Drivers Chapter 3 ...
Page 196: ...Bed Control Board PLD and Connector Chapter 3 ...
Page 197: ...Bed Control Board Motor Current Limiter Chapter 3 ...