en
Rev. 03 / 02.2021
AB3200
66 / 88
Software description
-] Settings
PROG
T/°C
RCF
RPM
t/min
Cool acc time = 15 s
PROG
T/°C
RCF
RPM
t/min
Cool acc time = 20 s
PROG
T/°C
RCF
RPM
t/min
Store Settings...
PROG
T/°C
RCF
RPM
t/min
-] Settings
PROG
T/°C
RCF
RPM
t/min
Cool dec speed = 150 rpm
PROG
T/°C
RCF
RPM
t/min
Cool dec speed = 100 rpm
PROG
T/°C
RCF
RPM
t/min
Store Settings...
PROG
T/°C
RCF
RPM
t/min
1.
Change to the menu
‘Machine Menu
🠂
Settings’
2.
Press the
[PROG] key until the ‘Cool acc time’ parameter is displayed.
3.
Use the
[adjusting knob] to set the desired value.
‘0’ = no delay time
4.
Press the
[START] key to save the setting.
5.
[STOP/OPEN] key Press 1x
⇒
Back to the menu
‘Machine Menu’.
6.
[STOP/OPEN] key Press 1x
⇒
Back to the main display
7.3.5.11
Cool dec speed
If required, it can be set that the cooling is no longer acti-
vated at the end of the centrifugation run during the run-
down, after reaching a set speed. This way, the stirring up
of sediments in the sample can be prevented. This speed
can be set from ‘0’ RPM to the maximum rotor speed
(Nmax) in increments of 10.
The speed can be set as follows:
1.
Change to the menu
‘Machine Menu
🠂
Settings’
2.
Press the
[PROG] key until the ‘Cool dec speed’ parameter is dis-
played.
3.
Use the
[adjusting knob] to set the desired value.
4.
Press the
[START] key to save the setting.
5.
[STOP/OPEN] key Press 1x
⇒
Back to the menu
‘Machine Menu’.
6.
[STOP/OPEN] key Press 1x
⇒
Back to the main display
7.3.5.12
Dual time mode
Further information on page 55
If the ‘Dual time mode’ function is activated, the beginning
of the runtime timer can be defined