Operation
Fig. 16: Viewing window in the
lid (figure is an example)
NOTICE
What to do in the event of unbalance
The ZentriMix is equipped with an unbalance switch like
every commercially available laboratory centrifuge.
If strong vibrations should result from incorrect loading or
process-intrinsic operations which the unbalance switch still
tolerates, but still leads to increased wear, the process must
be ended by pressing the
[STOP/OPEN] key twice.
6.8
Emergency unlocking
Personnel:
■
Authorized skilled personnel with special instruction
In the event of a power failure, the lid of the device cannot
be opened. Emergency unlocking must be done by hand.
DANGER
Danger due to rotor moving with high kinetic energy!
There is a risk of injury if the lid is opened via the emergency
release during device operation.
−
Only open the device via the emergency release when
the device is switched off.
−
Before the emergency release, make sure that the
device rotor has stopped.
To release the lid in an emergency, proceed as follows:
1.
Switch off the device and pull out the mains plug.
2.
Make sure the rotor is stopped by looking through the viewing
window in the lid.