Operation
6.5.3
Handling of bio-safety systems
Bio-safety systems may only be used when they are dry.
Each time before using the bio-safety system, all parts of
the bio-safety system must be visually inspected for
damage. In addition, the sealing ring(s) of the bio-safety
system must be checked to make sure they are in the cor-
rect installation position.
Damaged parts of the bio-safety system must be
exchanged immediately.
Damaged bio-safety systems are no longer microbiologi-
cally sealed.
To ensure leak-tightness, the lid of the bio-safety system
must be correctly sealed.
Available bio-safety systems can be found in the attach-
ment
Appendix A ‘Rotors and accessories’ on page 86.
If in doubt, you can get the information you need from the
manufacturer.
6.5.3.1
Opening and closing the lid of bio-safety systems
6.5.3.1.1
Lid with screw closure
To close the lid, proceed as follows:
1.
Place the lid on the hanger.
2.
Tightly seal the lid by hand by turning it clockwise.
To open the lid, proceed as follows:
Turn the lid by hand counterclockwise until the lid can be removed
from the hanger.
6.5.3.2
Storage of bio-safety systems
To prevent damage to the sealing rings during storage, bio-
safety systems may only be stored with the lid open.