47 / 88
en
Rev. 03 / 02.2021
AB3200
Software description
A 19 9/;\9 > 63< 2:00
PROG
T/°C
RCF
RPM
t/min
STO A
PROG
T/°C
RCF
RPM
t/min
STO D
PROG
T/°C
RCF
RPM
t/min
Multi program store..
PROG
T/°C
RCF
RPM
t/min
D 19 9/;\9 > 63< 2:00
PROG
T/°C
RCF
RPM
t/min
7.1.5
Saving a program link (STO A-Z)
In order to save the current program link (e.g. as a copy), proceed as fol-
lows:
1.
Press the
[PROG] key until ‘STO A’ is displayed.
2.
Use the
[adjusting knob] to set the desired program location.
3.
Press the
[START] key.
4.
You can now run the saved program link with the
[START] key with
the lid closed.
7.2
Centrifugation parameters
1 0 9/;\9 500 1:00
Fig. 25: Centrifugation parameters
If no key is pressed for 8 seconds after selection or during
parameter input, the previous values are shown on the dis-
play. The parameters must then be input again.
If multiple parameters are entered, the [START] key may
only be pressed after setting the last parameter in order to
accept all changes.
The input of parameters can be aborted at any time by
pressing the [STOP/OPEN] key. In this case, the settings
are not stored.
If parameters are changed, the program location number is
displayed in parentheses ‘( )’. This means that the centrifu-
gation data on the display no longer agrees with the stored
centrifugation data of the program location.